| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.029 | 1.939 |
| 25.00 | 1.660 | 1.570 |
| 12.50 | 1.185 | 1.095 |
| 6.25 | 0.813 | 0.723 |
| 3.13 | 0.501 | 0.411 |
| 1.56 | 0.343 | 0.253 |
| 0.78 | 0.172 | 0.082 |
| 0.00 | 0.090 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-114% | 99% |
| EDTA plasma(n=5) | 92-113% | 103% |
| Heparin plasma(n=5) | 93-102% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-115% | 89-112% | 84-110% | 82-92% |
| EDTA plasma(n=5) | 92-115% | 92-114% | 84-99% | 92-109% |
| Heparin plasma(n=5) | 86-92% | 81-102% | 94-104% | 94-102% |