| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.146 | 2.045 |
| 500.00 | 1.542 | 1.441 |
| 250.00 | 1.133 | 1.032 |
| 125.00 | 0.975 | 0.874 |
| 62.50 | 0.518 | 0.417 |
| 31.25 | 0.386 | 0.285 |
| 15.63 | 0.252 | 0.151 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 85-95% | 90% |
| Heparin plasma(n=5) | 79-93% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |
| EDTA plasma(n=5) | 86-93% | 91-105% | 87-98% | 89-94% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |