| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.112 | 2.015 |
| 25.00 | 1.649 | 1.552 |
| 12.50 | 1.193 | 1.096 |
| 6.25 | 0.930 | 0.833 |
| 3.13 | 0.516 | 0.419 |
| 1.56 | 0.359 | 0.262 |
| 0.78 | 0.171 | 0.074 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-114% | 98% |
| EDTA plasma(n=5) | 93-99% | 96% |
| Heparin plasma(n=5) | 86-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-99% | 90-95% | 87-93% | 93-109% |
| EDTA plasma(n=5) | 86-110% | 90-113% | 86-112% | 89-99% |
| Heparin plasma(n=5) | 85-99% | 82-113% | 80-89% | 95-106% |