| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.180 | 2.087 |
| 10.00 | 1.600 | 1.507 |
| 5.00 | 1.042 | 0.949 |
| 2.50 | 0.851 | 0.758 |
| 1.25 | 0.511 | 0.418 |
| 0.63 | 0.356 | 0.263 |
| 0.31 | 0.254 | 0.161 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-91% | 87% |
| EDTA plasma(n=5) | 85-109% | 97% |
| Heparin plasma(n=5) | 86-103% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-113% | 91-116% | 83-101% | 88-110% |
| EDTA plasma(n=5) | 89-107% | 83-113% | 82-96% | 87-104% |
| Heparin plasma(n=5) | 93-111% | 85-115% | 86-118% | 86-108% |