| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.063 | 1.983 |
| 5.00 | 1.633 | 1.553 |
| 2.50 | 1.250 | 1.170 |
| 1.25 | 0.811 | 0.731 |
| 0.63 | 0.550 | 0.470 |
| 0.31 | 0.256 | 0.176 |
| 0.16 | 0.177 | 0.097 |
| 0.00 | 0.080 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-96% | 91% |
| EDTA plasma(n=5) | 83-111% | 97% |
| Heparin plasma(n=5) | 81-110% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-112% | 83-97% | 92-100% | 85-108% |
| EDTA plasma(n=5) | 83-109% | 89-102% | 81-92% | 86-106% |
| Heparin plasma(n=5) | 86-98% | 82-100% | 84-89% | 88-94% |