| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.169 | 2.078 |
| 50.00 | 1.599 | 1.508 |
| 25.00 | 1.185 | 1.094 |
| 12.50 | 0.871 | 0.780 |
| 6.25 | 0.538 | 0.447 |
| 3.13 | 0.356 | 0.265 |
| 1.56 | 0.187 | 0.096 |
| 0.00 | 0.091 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-113% | 97% |
| EDTA plasma(n=5) | 88-121% | 105% |
| Heparin plasma(n=5) | 94-111% | 103% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-93% | 92-110% | 91-106% | 88-97% |
| EDTA plasma(n=5) | 83-94% | 81-98% | 95-101% | 86-96% |
| Heparin plasma(n=5) | 89-119% | 90-114% | 89-100% | 87-109% |