| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.204 | 2.101 |
| 25.00 | 1.527 | 1.424 |
| 12.50 | 1.142 | 1.039 |
| 6.25 | 0.752 | 0.649 |
| 3.13 | 0.535 | 0.432 |
| 1.57 | 0.389 | 0.286 |
| 0.79 | 0.225 | 0.122 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-107% | 100% |
| EDTA plasma(n=5) | 94-105% | 98% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-93% | 91-103% | 82-99% | 99-104% |
| EDTA plasma(n=5) | 86-93% | 91-105% | 87-98% | 89-94% |
| Heparin plasma(n=5) | 88-95% | 98-107% | 90-99% | 88-97% |