| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.093 | 1.984 |
| 25.00 | 1.683 | 1.574 |
| 12.50 | 1.298 | 1.189 |
| 6.25 | 0.803 | 0.694 |
| 3.13 | 0.564 | 0.455 |
| 1.56 | 0.381 | 0.272 |
| 0.78 | 0.232 | 0.123 |
| 0.00 | 0.109 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-118% | 100% |
| EDTA plasma(n=5) | 94-103% | 99% |
| Heparin plasma(n=5) | 90-107% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-91% | 82-119% | 84-100% | 80-113% |
| EDTA plasma(n=5) | 88-115% | 84-118% | 90-110% | 94-102% |
| Heparin plasma(n=5) | 88-97% | 93-99% | 94-99% | 91-111% |