| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.213 | 2.112 |
| 500.00 | 1.365 | 1.264 |
| 250.00 | 1.075 | 0.974 |
| 125.00 | 0.748 | 0.647 |
| 62.50 | 0.573 | 0.472 |
| 31.25 | 0.383 | 0.282 |
| 15.63 | 0.224 | 0.123 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-94% | 90% |
| EDTA plasma(n=5) | 89-109% | 99% |
| Heparin plasma(n=5) | 82-92% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-107% | 95-102% | 82-118% | 90-103% |
| EDTA plasma(n=5) | 86-105% | 90-103% | 91-104% | 89-101% |
| Heparin plasma(n=5) | 93-103% | 94-104% | 93-99% | 94-105% |
