| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.196 | 2.106 |
| 5.00 | 1.618 | 1.528 |
| 2.50 | 1.058 | 0.968 |
| 1.25 | 0.732 | 0.642 |
| 0.63 | 0.485 | 0.395 |
| 0.31 | 0.322 | 0.232 |
| 0.16 | 0.202 | 0.112 |
| 0.00 | 0.090 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-101% | 95% |
| EDTA plasma(n=5) | 86-98% | 92% |
| Heparin plasma(n=5) | 94-101% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 94-112% | 90-102% | 82-118% | 83-100% |
| EDTA plasma(n=5) | 89-105% | 80-95% | 91-111% | 83-90% |
| Heparin plasma(n=5) | 85-108% | 81-111% | 84-96% | 90-109% |