| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.135 | 2.046 |
| 5.00 | 1.562 | 1.473 |
| 2.50 | 1.135 | 1.046 |
| 1.25 | 0.847 | 0.758 |
| 0.63 | 0.457 | 0.368 |
| 0.32 | 0.302 | 0.213 |
| 0.16 | 0.196 | 0.107 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 78-95% | 86% |
| Heparin plasma(n=5) | 80-96% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-98% | 88-104% | 86-102% | 79-93% |
| EDTA plasma(n=5) | 89-97% | 85-94% | 87-101% | 85-96% |
| Heparin plasma(n=5) | 85-94% | 87-101% | 82-95% | 82-93% |