| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.260 | 2.192 |
| 1000.00 | 1.501 | 1.433 |
| 500.00 | 0.864 | 0.796 |
| 250.00 | 0.509 | 0.441 |
| 125.00 | 0.258 | 0.190 |
| 62.50 | 0.160 | 0.092 |
| 31.25 | 0.119 | 0.051 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-113% | 97% |
| EDTA plasma(n=5) | 81-100% | 91% |
| Heparin plasma(n=5) | 95-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-103% | 94-105% | 92-118% | 92-114% |
| EDTA plasma(n=5) | 94-104% | 93-99% | 91-119% | 81-112% |
| Heparin plasma(n=5) | 83-94% | 85-117% | 86-108% | 89-101% |