| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.155 | 2.025 |
| 500.00 | 1.460 | 1.330 |
| 250.00 | 0.886 | 0.756 |
| 125.00 | 0.536 | 0.406 |
| 62.50 | 0.301 | 0.171 |
| 31.25 | 0.245 | 0.115 |
| 15.63 | 0.223 | 0.093 |
| 0.00 | 0.130 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-117% | 103% |
| EDTA plasma(n=5) | 92-108% | 100% |
| Heparin plasma(n=5) | 82-94% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-94% | 95-117% | 88-95% | 92-119% |
| EDTA plasma(n=5) | 88-109% | 91-114% | 89-118% | 83-89% |
| Heparin plasma(n=5) | 94-110% | 89-105% | 81-88% | 82-93% |