| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 6000.00 | 2.328 | 2.245 |
| 3000.00 | 1.645 | 1.562 |
| 1500.00 | 1.176 | 1.093 |
| 750.00 | 0.892 | 0.809 |
| 375.00 | 0.547 | 0.464 |
| 187.50 | 0.336 | 0.253 |
| 93.75 | 0.252 | 0.169 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 98-106% | 87-98% | 91-102% |
| EDTA plasma(n=5) | 79-92% | 85-96% | 80-98% | 95-102% |
| Heparin plasma(n=5) | 89-99% | 83-95% | 98-104% | 96-106% |