| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.079 | 1.993 |
| 25.00 | 1.556 | 1.470 |
| 12.50 | 1.121 | 1.035 |
| 6.25 | 0.975 | 0.889 |
| 3.13 | 0.556 | 0.470 |
| 1.57 | 0.305 | 0.219 |
| 0.79 | 0.187 | 0.101 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 80-92% | 86% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-97% | 87-96% | 85-96% |
| EDTA plasma(n=5) | 95-103% | 89-101% | 87-98% | 85-92% |
| Heparin plasma(n=5) | 98-103% | 86-92% | 87-93% | 93-102% |