| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.135 | 2.039 |
| 25.00 | 1.562 | 1.466 |
| 12.50 | 1.169 | 1.073 |
| 6.25 | 0.897 | 0.801 |
| 3.13 | 0.556 | 0.460 |
| 1.57 | 0.305 | 0.209 |
| 0.79 | 0.208 | 0.112 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-107% | 100% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 85-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-92% | 85-96% | 80-98% | 95-102% |
| EDTA plasma(n=5) | 89-99% | 83-95% | 98-104% | 96-106% |
| Heparin plasma(n=5) | 85-100% | 84-98% | 95-101% | 88-102% |