Products

Human TSP-1(Thrombospondin-1) ELISA Kit
One-click to copy product information
ELK10311
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human TSP-1(Thrombospondin-1) ELISA Kit
Reactivity: Human
Alternative Names: TSP1; Thrombospondin-1p180; THBS1; Thrombospondin 1
Assay Type: Sandwich
Sensitivity: 0.47 ng/mL
Standard: 100 ng/mL
Detection Range: 1.56-100 ng/mL
Sample type: plasma
Assay length: 3.5h
Research Area: Infection immunity;Hematology;
Uniprot ID: P07996
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human TSP-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human TSP-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human TSP-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human TSP-1 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
100.00 2.164 2.040
50.00 1.782 1.658
25.00 1.205 1.081
12.50 0.963 0.839
6.25 0.547 0.423
3.13 0.339 0.215
1.57 0.227 0.103
0.00 0.124 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant TSP-1 and the recovery rates were calculated by comparing the measured value to the expected amount of TSP-1 in samples.
Matrix Recovery range Average
serum(n=5) 85-99% 92%
EDTA plasma(n=5) 93-107% 100%
Heparin plasma(n=5) 87-99% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TSP-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-97% 93-105% 82-93% 95-104%
EDTA plasma(n=5) 87-96% 85-97% 85-94% 87-96%
Heparin plasma(n=5) 93-104% 83-96% 95-103% 88-95%