| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.019 | 1.938 |
| 250.00 | 1.421 | 1.340 |
| 125.00 | 0.817 | 0.736 |
| 62.50 | 0.587 | 0.506 |
| 31.25 | 0.353 | 0.272 |
| 15.63 | 0.257 | 0.176 |
| 7.81 | 0.111 | 0.030 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-111% | 101% |
| EDTA plasma(n=5) | 85-116% | 101% |
| Heparin plasma(n=5) | 82-117% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-114% | 83-105% | 88-115% | 82-111% |
| EDTA plasma(n=5) | 85-105% | 92-103% | 87-96% | 82-93% |
| Heparin plasma(n=5) | 83-112% | 83-96% | 82-93% | 94-114% |