| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.249 | 2.156 |
| 1000.00 | 1.700 | 1.607 |
| 500.00 | 1.204 | 1.111 |
| 250.00 | 0.919 | 0.826 |
| 125.00 | 0.565 | 0.472 |
| 62.50 | 0.320 | 0.227 |
| 31.25 | 0.197 | 0.104 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-102% | 97% |
| EDTA plasma(n=5) | 89-108% | 99% |
| Heparin plasma(n=5) | 88-110% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-102% | 90-109% | 88-113% | 82-90% |
| EDTA plasma(n=5) | 88-112% | 90-100% | 91-114% | 86-115% |
| Heparin plasma(n=5) | 82-104% | 81-92% | 92-99% | 83-95% |