| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.360 | 2.273 |
| 500.00 | 1.600 | 1.513 |
| 250.00 | 0.916 | 0.829 |
| 125.00 | 0.511 | 0.424 |
| 62.50 | 0.301 | 0.214 |
| 31.25 | 0.177 | 0.090 |
| 15.63 | 0.115 | 0.028 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-88% | 85% |
| EDTA plasma(n=5) | 93-125% | 109% |
| Heparin plasma(n=5) | 82-115% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-104% | 86-105% | 82-120% | 92-112% |
| EDTA plasma(n=5) | 87-96% | 83-110% | 92-116% | 92-104% |
| Heparin plasma(n=5) | 89-95% | 91-116% | 88-102% | 83-106% |