| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.940 | 1.867 |
| 500.00 | 1.225 | 1.152 |
| 250.00 | 0.662 | 0.589 |
| 125.00 | 0.356 | 0.283 |
| 62.50 | 0.170 | 0.097 |
| 31.25 | 0.177 | 0.104 |
| 15.63 | 0.154 | 0.081 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-95% | 90% |
| EDTA plasma(n=5) | 93-100% | 97% |
| Heparin plasma(n=5) | 84-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-98% | 93-107% | 94-101% | 80-102% |
| EDTA plasma(n=5) | 92-119% | 88-110% | 82-95% | 86-104% |
| Heparin plasma(n=5) | 84-93% | 87-102% | 94-106% | 90-99% |