| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.087 | 2.007 |
| 2500.00 | 1.671 | 1.591 |
| 1250.00 | 1.179 | 1.099 |
| 625.00 | 0.841 | 0.761 |
| 312.50 | 0.615 | 0.535 |
| 156.25 | 0.436 | 0.356 |
| 78.13 | 0.217 | 0.137 |
| 0.00 | 0.080 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-105% | 96% |
| EDTA plasma(n=5) | 89-113% | 101% |
| Heparin plasma(n=5) | 86-115% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-113% | 85-97% | 92-115% | 92-99% |
| EDTA plasma(n=5) | 90-97% | 88-94% | 95-116% | 87-115% |
| Heparin plasma(n=5) | 80-93% | 83-95% | 95-108% | 86-111% |