| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.941 | 1.870 |
| 50.00 | 1.509 | 1.438 |
| 25.00 | 1.069 | 0.998 |
| 12.50 | 0.749 | 0.678 |
| 6.25 | 0.581 | 0.510 |
| 3.13 | 0.315 | 0.244 |
| 1.57 | 0.152 | 0.081 |
| 0.00 | 0.071 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 88-102% | 95% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 86-92% | 81-93% | 89-105% |
| EDTA plasma(n=5) | 85-94% | 79-96% | 92-101% | 83-90% |
| Heparin plasma(n=5) | 95-102% | 93-101% | 97-105% | 95-103% |