| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.827 | 1.744 |
| 250.00 | 1.483 | 1.400 |
| 125.00 | 0.927 | 0.844 |
| 62.50 | 0.602 | 0.519 |
| 31.25 | 0.456 | 0.373 |
| 15.63 | 0.314 | 0.231 |
| 7.81 | 0.209 | 0.126 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-117% | 105% |
| EDTA plasma(n=5) | 92-102% | 97% |
| Heparin plasma(n=5) | 93-119% | 106% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-117% | 89-110% | 87-108% | 88-105% |
| EDTA plasma(n=5) | 82-114% | 87-115% | 84-90% | 85-94% |
| Heparin plasma(n=5) | 82-118% | 92-98% | 84-105% | 95-116% |