| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.842 | 1.759 |
| 250.00 | 1.502 | 1.419 |
| 125.00 | 0.939 | 0.856 |
| 62.50 | 0.615 | 0.532 |
| 31.25 | 0.451 | 0.368 |
| 15.63 | 0.312 | 0.229 |
| 7.82 | 0.214 | 0.131 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-105% | 98% |
| EDTA plasma(n=5) | 84-96% | 90% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-93% | 91-105% | 87-98% | 89-94% |
| EDTA plasma(n=5) | 93-102% | 93-106% | 87-101% | 91-104% |
| Heparin plasma(n=5) | 89-97% | 93-102% | 81-96% | 93-106% |