| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 25.00 | 2.193 | 2.112 |
| 12.50 | 1.819 | 1.738 |
| 6.25 | 1.204 | 1.123 |
| 3.13 | 0.742 | 0.661 |
| 1.56 | 0.556 | 0.475 |
| 0.78 | 0.350 | 0.269 |
| 0.39 | 0.219 | 0.138 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-93% | 87% |
| EDTA plasma(n=5) | 88-94% | 91% |
| Heparin plasma(n=5) | 87-115% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-118% | 83-120% | 93-106% | 83-104% |
| EDTA plasma(n=5) | 94-104% | 91-109% | 91-110% | 93-116% |
| Heparin plasma(n=5) | 94-117% | 86-94% | 86-96% | 80-97% |