| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.929 | 1.849 |
| 50.00 | 1.605 | 1.525 |
| 25.00 | 1.051 | 0.971 |
| 12.50 | 0.767 | 0.687 |
| 6.25 | 0.496 | 0.416 |
| 3.13 | 0.369 | 0.289 |
| 1.56 | 0.203 | 0.123 |
| 0.00 | 0.080 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-104% | 96% |
| EDTA plasma(n=5) | 87-105% | 96% |
| Heparin plasma(n=5) | 91-114% | 103% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-115% | 91-100% | 86-116% | 88-97% |
| EDTA plasma(n=5) | 92-117% | 89-111% | 84-113% | 88-99% |
| Heparin plasma(n=5) | 92-109% | 86-94% | 86-113% | 84-118% |