| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 150.00 | 1.941 | 1.838 |
| 75.00 | 1.484 | 1.381 |
| 37.50 | 1.069 | 0.966 |
| 18.75 | 0.749 | 0.646 |
| 9.38 | 0.581 | 0.478 |
| 4.69 | 0.415 | 0.312 |
| 2.35 | 0.241 | 0.138 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-103% | 95% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-104% | 90-102% | 87-98% | 95-103% |
| EDTA plasma(n=5) | 88-97% | 96-102% | 81-93% | 89-97% |
| Heparin plasma(n=5) | 85-94% | 87-96% | 82-95% | 93-102% |