| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.770 | 1.674 |
| 50.00 | 1.285 | 1.189 |
| 25.00 | 0.810 | 0.714 |
| 12.50 | 0.561 | 0.465 |
| 6.25 | 0.415 | 0.319 |
| 3.13 | 0.380 | 0.284 |
| 1.56 | 0.211 | 0.115 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 84-112% | 98% |
| Heparin plasma(n=5) | 90-117% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-103% | 93-107% | 81-94% | 92-104% |
| EDTA plasma(n=5) | 83-104% | 95-113% | 89-120% | 92-119% |
| Heparin plasma(n=5) | 83-94% | 82-100% | 91-109% | 83-103% |