| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.871 | 1.784 |
| 1000.00 | 1.549 | 1.462 |
| 500.00 | 1.023 | 0.936 |
| 250.00 | 0.685 | 0.598 |
| 125.00 | 0.427 | 0.340 |
| 62.50 | 0.307 | 0.220 |
| 31.25 | 0.210 | 0.123 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-90% | 88% |
| EDTA plasma(n=5) | 82-107% | 95% |
| Heparin plasma(n=5) | 89-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-109% | 89-100% | 94-111% | 80-87% |
| EDTA plasma(n=5) | 85-111% | 95-102% | 83-105% | 87-104% |
| Heparin plasma(n=5) | 85-115% | 85-112% | 86-95% | 85-109% |