| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.846 | 1.760 |
| 1000.00 | 1.527 | 1.441 |
| 500.00 | 1.034 | 0.948 |
| 250.00 | 0.675 | 0.589 |
| 125.00 | 0.424 | 0.338 |
| 62.50 | 0.312 | 0.226 |
| 31.25 | 0.208 | 0.122 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-98% | 92% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 97-103% | 87-98% | 95-102% |
| EDTA plasma(n=5) | 98-103% | 86-92% | 87-93% | 93-102% |
| Heparin plasma(n=5) | 85-94% | 97-106% | 92-101% | 82-90% |