| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.946 | 1.858 |
| 10.00 | 1.686 | 1.598 |
| 5.00 | 1.205 | 1.117 |
| 2.50 | 0.871 | 0.783 |
| 1.25 | 0.577 | 0.489 |
| 0.63 | 0.357 | 0.269 |
| 0.32 | 0.238 | 0.150 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-95% | 86% |
| EDTA plasma(n=5) | 84-98% | 91% |
| Heparin plasma(n=5) | 90-105% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-101% | 82-98% | 89-102% |
| EDTA plasma(n=5) | 91-97% | 81-93% | 89-103% | 82-91% |
| Heparin plasma(n=5) | 92-101% | 80-94% | 86-101% | 89-98% |