| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.136 | 2.041 |
| 50.00 | 1.629 | 1.534 |
| 25.00 | 1.182 | 1.087 |
| 12.50 | 0.831 | 0.736 |
| 6.25 | 0.582 | 0.487 |
| 3.13 | 0.363 | 0.268 |
| 1.57 | 0.258 | 0.163 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-93% | 85% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 98-105% | 86-95% | 94-103% | 95-102% |
| EDTA plasma(n=5) | 87-101% | 95-103% | 97-105% | 88-97% |
| Heparin plasma(n=5) | 83-92% | 92-96% | 86-98% | 95-101% |