| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.908 | 1.820 |
| 5.00 | 1.647 | 1.559 |
| 2.50 | 1.182 | 1.094 |
| 1.25 | 0.805 | 0.717 |
| 0.63 | 0.543 | 0.455 |
| 0.32 | 0.324 | 0.236 |
| 0.16 | 0.215 | 0.127 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-106% | 100% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 77-95% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-106% | 88-103% | 81-98% | 93-102% |
| EDTA plasma(n=5) | 85-96% | 86-98% | 92-103% | 87-93% |
| Heparin plasma(n=5) | 81-94% | 85-97% | 93-102% | 81-95% |