| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.186 | 2.085 |
| 1000.00 | 1.702 | 1.601 |
| 500.00 | 1.166 | 1.065 |
| 250.00 | 0.979 | 0.878 |
| 125.00 | 0.590 | 0.489 |
| 62.50 | 0.325 | 0.224 |
| 31.25 | 0.211 | 0.110 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-99% | 91% |
| EDTA plasma(n=5) | 91-118% | 105% |
| Heparin plasma(n=5) | 91-112% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-103% | 87-97% | 89-94% | 88-110% |
| EDTA plasma(n=5) | 86-109% | 88-108% | 91-101% | 92-112% |
| Heparin plasma(n=5) | 95-107% | 86-116% | 93-113% | 90-118% |