| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.266 | 2.170 |
| 5.00 | 1.599 | 1.503 |
| 2.50 | 1.132 | 1.036 |
| 1.25 | 0.847 | 0.751 |
| 0.63 | 0.567 | 0.471 |
| 0.32 | 0.369 | 0.273 |
| 0.16 | 0.226 | 0.130 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 90-107% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-93% | 87-101% | 96-105% | 96-105% |
| EDTA plasma(n=5) | 99-106% | 78-92% | 87-95% | 82-94% |
| Heparin plasma(n=5) | 92-105% | 83-92% | 85-96% | 78-90% |