| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.266 | 2.170 | 
| 500.00 | 1.576 | 1.480 | 
| 250.00 | 1.286 | 1.190 | 
| 125.00 | 0.907 | 0.811 | 
| 62.50 | 0.545 | 0.449 | 
| 31.25 | 0.369 | 0.273 | 
| 15.63 | 0.215 | 0.119 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-97% | 90% | 
| EDTA plasma(n=5) | 92-107% | 99% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 81-93% | 86-98% | 91-103% | 89-101% | 
| EDTA plasma(n=5) | 83-96% | 78-98% | 88-97% | 85-92% | 
| Heparin plasma(n=5) | 86-95% | 94-103% | 95-102% | 87-101% |