48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.075 | 1.993 |
500.00 | 1.663 | 1.581 |
250.00 | 1.196 | 1.114 |
125.00 | 0.834 | 0.752 |
62.50 | 0.563 | 0.481 |
31.25 | 0.361 | 0.279 |
15.63 | 0.230 | 0.148 |
0.00 | 0.082 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-95% | 86% |
EDTA plasma(n=5) | 89-105% | 97% |
Heparin plasma(n=5) | 86-99% | 92% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-94% | 79-96% | 85-99% | 80-92% |
EDTA plasma(n=5) | 84-100% | 95-103% | 88-102% | 91-99% |
Heparin plasma(n=5) | 84-98% | 89-102% | 90-99% | 87-103% |