| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.154 | 2.069 |
| 5.00 | 1.708 | 1.623 |
| 2.50 | 1.223 | 1.138 |
| 1.25 | 0.975 | 0.890 |
| 0.63 | 0.585 | 0.500 |
| 0.32 | 0.366 | 0.281 |
| 0.16 | 0.212 | 0.127 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-105% | 98% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 82-98% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-101% | 86-97% | 88-95% | 85-97% |
| EDTA plasma(n=5) | 80-92% | 86-102% | 98-105% | 86-92% |
| Heparin plasma(n=5) | 96-105% | 89-103% | 85-94% | 87-96% |