| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.387 | 2.290 |
| 5.00 | 1.655 | 1.558 |
| 2.50 | 1.167 | 1.070 |
| 1.25 | 0.921 | 0.824 |
| 0.63 | 0.607 | 0.510 |
| 0.31 | 0.366 | 0.269 |
| 0.16 | 0.220 | 0.123 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-88% | 85% |
| EDTA plasma(n=5) | 92-98% | 95% |
| Heparin plasma(n=5) | 90-110% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-118% | 84-101% | 86-111% | 93-115% |
| EDTA plasma(n=5) | 80-93% | 83-102% | 92-104% | 81-90% |
| Heparin plasma(n=5) | 83-94% | 83-98% | 93-114% | 93-108% |