| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.354 | 2.258 |
| 5.00 | 1.668 | 1.572 |
| 2.50 | 1.135 | 1.039 |
| 1.25 | 0.897 | 0.801 |
| 0.63 | 0.614 | 0.518 |
| 0.32 | 0.357 | 0.261 |
| 0.16 | 0.216 | 0.120 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 95-107% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-100% | 84-99% | 85-97% | 86-97% |
| EDTA plasma(n=5) | 86-92% | 89-103% | 89-98% | 79-85% |
| Heparin plasma(n=5) | 81-91% | 93-106% | 89-97% | 85-94% |