| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.266 | 2.178 | 
| 5.00 | 1.694 | 1.606 | 
| 2.50 | 1.121 | 1.033 | 
| 1.25 | 0.875 | 0.787 | 
| 0.63 | 0.572 | 0.484 | 
| 0.32 | 0.323 | 0.235 | 
| 0.16 | 0.192 | 0.104 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-106% | 99% | 
| EDTA plasma(n=5) | 81-95% | 88% | 
| Heparin plasma(n=5) | 90-107% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-99% | 85-96% | 81-95% | 95-105% | 
| EDTA plasma(n=5) | 88-102% | 93-101% | 90-101% | 87-98% | 
| Heparin plasma(n=5) | 85-99% | 98-105% | 92-101% | 89-105% |