| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.154 | 2.081 |
| 100.00 | 1.642 | 1.569 |
| 50.00 | 1.269 | 1.196 |
| 25.00 | 0.963 | 0.890 |
| 12.50 | 0.487 | 0.414 |
| 6.25 | 0.318 | 0.245 |
| 3.13 | 0.208 | 0.135 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-106% | 100% |
| EDTA plasma(n=5) | 78-96% | 87% |
| Heparin plasma(n=5) | 89-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-101% | 93-104% | 81-99% | 94-106% |
| EDTA plasma(n=5) | 83-97% | 88-96% | 83-96% | 78-98% |
| Heparin plasma(n=5) | 92-101% | 97-105% | 84-96% | 86-97% |