| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.994 | 1.899 |
| 2500.00 | 1.562 | 1.467 |
| 1250.00 | 1.088 | 0.993 |
| 625.00 | 0.796 | 0.701 |
| 312.50 | 0.508 | 0.413 |
| 156.25 | 0.305 | 0.210 |
| 78.13 | 0.175 | 0.080 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 80-94% | 87% |
| Heparin plasma(n=5) | 90-107% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-101% | 88-96% |
| EDTA plasma(n=5) | 86-98% | 91-103% | 82-95% | 87-95% |
| Heparin plasma(n=5) | 85-96% | 86-97% | 87-102% | 89-101% |