| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.045 | 1.952 | 
| 1000.00 | 1.708 | 1.615 | 
| 500.00 | 1.193 | 1.100 | 
| 250.00 | 0.933 | 0.840 | 
| 125.00 | 0.508 | 0.415 | 
| 62.50 | 0.354 | 0.261 | 
| 31.25 | 0.236 | 0.143 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-106% | 97% | 
| EDTA plasma(n=5) | 84-96% | 90% | 
| Heparin plasma(n=5) | 87-102% | 94% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-96% | 93-101% | 95-103% | 95-103% | 
| EDTA plasma(n=5) | 82-96% | 86-98% | 85-98% | 95-104% | 
| Heparin plasma(n=5) | 83-97% | 87-105% | 87-99% | 89-95% |