| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.191 | 2.109 |
| 2500.00 | 1.644 | 1.562 |
| 1250.00 | 1.020 | 0.938 |
| 625.00 | 0.862 | 0.780 |
| 312.50 | 0.563 | 0.481 |
| 156.25 | 0.297 | 0.215 |
| 78.13 | 0.211 | 0.129 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-108% | 96% |
| EDTA plasma(n=5) | 90-129% | 110% |
| Heparin plasma(n=5) | 88-99% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-105% | 81-102% | 87-93% | 86-112% |
| EDTA plasma(n=5) | 81-107% | 91-112% | 91-117% | 95-112% |
| Heparin plasma(n=5) | 90-119% | 85-99% | 85-111% | 84-101% |