| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 0.189 | |
| 5.00 | 0.386 | |
| 2.50 | 0.596 | |
| 1.25 | 0.895 | |
| 0.63 | 1.253 | |
| 0.32 | 1.565 | |
| 0.16 | 1.761 | |
| 0.00 | 2.101 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-99% | 92% |
| EDTA plasma(n=5) | 89-103% | 96% |
| Heparin plasma(n=5) | 78-96% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 79-92% | 82-95% | 86-97% |
| EDTA plasma(n=5) | 85-92% | 89-103% | 97-105% | 91-99% |
| Heparin plasma(n=5) | 86-97% | 90-98% | 80-93% | 92-105% |