| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 0.227 | |
| 10.00 | 0.397 | |
| 5.00 | 0.575 | |
| 2.50 | 0.833 | |
| 1.25 | 1.275 | |
| 0.63 | 1.578 | |
| 0.32 | 1.799 | |
| 0.00 | 2.112 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-98% | 79-93% | 92-101% | 82-90% |
| EDTA plasma(n=5) | 85-92% | 80-96% | 87-98% | 95-102% |
| Heparin plasma(n=5) | 88-103% | 83-94% | 92-105% | 89-97% |