Products

EasyStep Human hCG(Hyperglycosylated Chorionic Gonadotropin) ELISA Kit
One-click to copy product information
ELK089ES
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: EasyStep Human hCG(Hyperglycosylated Chorionic Gonadotropin) ELISA Kit
Reactivity: Human
Alternative Names: hCG; Chorionic Gonadotrophin
Assay Type: Sandwich
Sensitivity: 0.08 mIU/mL
Standard: 500mIU/mL
Detection Range: 7.81-500 mIU/mL
Sample type: Serum, plasma
Assay length: 1.5h
Research Area: Endocrinology;Reproductive science;Neuro science;Genetic science;Hormone metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human hCG, and the Human hCG standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human hCG. After TMB substrate solution is added, only those wells that contain Human hCG and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human hCG in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (mIU/mL) OD Corrected OD
500.00 2.126 2.044
250.00 1.587 1.505
125.00 1.255 1.173
62.50 0.821 0.739
31.25 0.553 0.471
15.63 0.256 0.174
7.82 0.175 0.093
0.00 0.082 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant hCG and the recovery rates were calculated by comparing the measured value to the expected amount of hCG in samples.
Matrix Recovery range Average
serum(n=5) 83-95% 89%
EDTA plasma(n=5) 86-99% 92%
Heparin plasma(n=5) 80-95% 87%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of hCG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-104% 82-93% 81-96% 83-94%
EDTA plasma(n=5) 80-92% 86-102% 87-103% 85-101%
Heparin plasma(n=5) 82-93% 96-105% 87-98% 85-97%

Product Components