| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 0.232 | |
| 100.00 | 0.336 | |
| 50.00 | 0.586 | |
| 25.00 | 0.895 | |
| 12.50 | 1.257 | |
| 6.25 | 1.596 | |
| 3.13 | 1.889 | |
| 0.00 | 2.232 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-103% | 96% |
| EDTA plasma(n=5) | 81-96% | 88% |
| Heparin plasma(n=5) | 78-94% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-95% | 89-102% | 85-92% | 96-104% |
| EDTA plasma(n=5) | 78-92% | 83-96% | 87-101% | 97-106% |
| Heparin plasma(n=5) | 86-97% | 95-101% | 88-102% | 81-93% |